Common Fault and Solution of Hydraulic Brinell Hardness Tester

Table of Content [Hide]
    Nov 22 , 2021

    The 5mm ball fell off in the test head


    The solution is as below:


    Remove the test head from the shelf, anticlockwise the screw, you can see the oil overflowed, draw out the oil with the needle and take out the spring, keep to draw out the oil until you can see the φ5mm ball groove, use the light to check if theφ5mm ball is moved and not in the ball groove, if the φ5mm ball is in working condition, normally if the pressure lever stick not working, theφ5mm ball is moved, shake the test head to let theφ5mm ball back to the ball groove, and then filled with the oil slowly to avoid entering the air, put the spring back, and clockwise the screw, fix the test head back on the shelf.


    Common Fault and Solution of Hydraulic Brinell Hardness Tester

                             Tools                                       φ5mm ball not in ball groove


    Joanna Zhao
    Joanna Zhao

    Hi, I'm Joanna Zhao, the Import & Export Manager of Shenyang TX Testing Instruments Inc. I've been with TX Testing for 15 years, gaining hands-on experience by closely observing the production process and personally assembling and testing our portable hardness testers. This has deepened my technical understanding and strengthened my confidence in the industry's development. With this foundation, I provide professional support, feasibility analyses, and customized solutions to meet diverse customer needs, offering one-stop services that ensure efficiency and satisfaction.

    Products
    References
    We use cookies to optimise and personalise your experience, but you can choose to opt out of non-essential cookies.
    To find out more, read our Privacy Policy
    Reject All
    Accept All